Title: | Simple method to reduce the dead time of multichannel analyzers | Authors: | Puzović, Jovan Aničin, I. V. |
Issue Date: | 1-Jan-1999 | Journal: | Review of Scientific Instruments | Abstract: | A method to reduce the dead time of multichannel analyzers which uses multiple sample-and-hold circuits in the input circuitry of the analyzer's analog-digital converter is proposed. Due to the exponential distribution of time intervals between successive pulses, the dead time is greatly reduced with even a small number of sample-and-hold circuits. The effects of the method are analyzed and discussed by means of Monte Carlo simulations. © 1999 American Institute of Physics. |
URI: | https://physrep.ff.bg.ac.rs/handle/123456789/882 | ISSN: | 0034-6748 | DOI: | 10.1063/1.1149814 |
Appears in Collections: | Journal Article |