Title: An off-line method for the correction of amplitude walk in leading-edge timing
Authors: Puzović, Jovan
Aničin, I. V.
Keywords: Amplitude walk;Coincidence system;Leading-edge timing;Multiparameter;Off-line correct
Issue Date: 11-Mar-2007
Journal: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Abstract: 
It is demonstrated that in the leading-edge timing of slow amplifier pulses the functional dependence of timing on amplitude, or amplitude walk, is regular and may be satisfactorily analytically described. In a multiparameter coincidence system working in the list mode this dependence can then be used to off-line correct for this systematic effect in the wide dynamic range. The quality of the correction is illustrated by the time-resolved measurement of the effects induced by neutrons propagating in different media. © 2007 Elsevier B.V. All rights reserved.