Please use this identifier to cite or link to this item: https://physrep.ff.bg.ac.rs/handle/123456789/885
Title: Towards the definition of the optimum height of voluminous cylindrical sources in gamma-ray spectroscopy
Authors: Vesić, D.
Puzović, Jovan
Aničin, I. V.
Keywords: Errors;Gamma-spectroscopy;Large sources;Optimum quantity
Issue Date: 1-Jan-1999
Journal: Applied Radiation and Isotopes
Abstract: 
We show that for cylindrical voluminous sources in gamma-ray spectroscopy the relative error of the spectral intensity of a low-energy spectral line, as the height of the source increases, saturates faster than the spectral intensity itself. This significantly reduces the quantity of the required sample material as compared to that based on the usual considerations of the spectral intensity only. We also demonstrate that there exists a weakly pronounced height of the source (the source size) which minimizes the relative error of the intensity of a given low-energy spectral line. Due to the smallness of the effect the benefit of introducing the concept of the truly optimum size of the source is more on a psychological side, one should feel easier using an optimum quantity of the source material than merely a sufficient one, as is otherwise the case.
URI: https://physrep.ff.bg.ac.rs/handle/123456789/885
ISSN: 0969-8043
DOI: 10.1016/S0969-8043(98)00172-9
Appears in Collections:Journal Article

Show full item record

Page view(s)

6
checked on Nov 20, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.